Labconco Class II, Type B2 Biological Safety Cabinets
Manufacturer: Labconco
Model: 3441000
Purifier Logic Class II, Type B2 Biosafety Cabinets provide personnel, product and environmental protection from hazardous particulates such as agents that require Biosafety Level 1, 2 or 3 containment. Other appropriate applications include work with antineoplastic drugs, genetic material, asbestos and additional substances that generate hazardous airborne particulates. All models are NSF-Listed.
These biological safety cabinets are designed to be hard ducted to the outside. During operation, room air is drawn into the top of the cabinet and through a supply HEPA filter. This filtered air flows downward through the work area. Room air is also drawn into the inlet grille located at the work access opening. All of the contaminated air passes through the exhaust HEPA filter. A dedicated exhaust system and remote blower discharge 100% of the filtered exhaust air from the laboratory. Since none of the air is recirculated, these biohazard cabinets may be used for work with agents treated with volatile toxic chemicals and radionuclides.
Features
• Nominal inflow velocity of 105 fpm (0.5 m/sec)
• Nominal downflow velocity of 55 fpm (0.3 m/sec)
• Patent-pending design
• Interior-mounted, line-of-sight LCD information center with "Filter Life Remaining" bar graph, status line for alarm conditions and alerts to warn when filter life diminishes to 20%, 10% and 0%
• Filter monitoring system consisting of an electronically commutated motor (ECM) that delivers a precise volume of air as required and automatically adjusts as filters load without relying on airflow sensors
• Smart-Start™ System that allows the user to program start up and shut down operations
• Night-Smart™ System that idles the blower when the sash is fully closed
• Built-in interval or elapsed timer for experiment monitoring or UV light control (on models with UV light)
• Touchpad control on right-hand side post for manual activation of blower, light, timer, audible alarm mute and menu selection
• Radiused type 304 stainless steel interior and removable, seamless, dished work surface with lift out knobs
• Epoxy-coated steel exterior
• Towel catch located under work surface
• Contain-Air™ Negative Pressure Channel
• Class 5 conditions per ISO 14644-1 and 2 (formerly Class 100)
• Supply and exhaust 99.99% efficient HEPA filters
• Two electrical duplex receptacles, (single outlets on 230 volt models), located one on each side, with ground fault interruption and stainless steel splash covers
• Fully-closing, clear 1/4" tempered safety glass sash with two sash handles; counterbalanced, anti-racking mechanism; and 10° slope
• 29" sash opening viewing height
• Curved stainless steel inlet grille with Reserve-Air™ Secondary Airflow Slots
• Bright, 100 foot-candle, glare-free fluorescent lighting located outside the contaminated work area
• Intrinsically safe negative pressure design
• Electronic security lock (optional activation) that requires code to operate the cabinet
• 10" diameter exhaust outlet
• Full three year warranty
• NSF-Listed
• ETL listed (115 volt, 60 Hz models)
• CE conformity (230 volt, 50 Hz models)
Compliance
• NSF International Standard 49 for Biohazard Cabinetry
• UL Standard 61010-1
• CAN/CSA Standard C22.2 No. 1010.1
• CE conformity (230 volt models)
Specifications